Helley-Smith Sampler
Helley-Smith Sampler
The Helley-Smith sampler is suitable for the determination of bed load transport, especially in streams or smaller rivers. The material goes through the 7.6 x 7.6 cm opening into a 250 micron mesh bag.
Example: The amount can then be determined as follows: Sb = Ks (1-p) V / (bT) or Sb = G / (bT)
k = calibration factor
p = porosity factor (= 0.4)
s = density of sediment particles (= 2650 kg / m 3)
V = volume G = dry matter (kg)
b = width of the inlet cross-section (= 0.0762 m)